e-ISSN: 2717-7157
  • Home
  • Aims and Scope
  • Editorial Board
  • Instructions for Authors
  • Table of Contents
  • Contact
  • Türkçe
Current Issue
Archive
Search
Online Subscription Login
Subscription Form
Most Popular
Download Articles Read Articles
Automated Perimetry
Carbonic Anhydrase Inhibitors
Intra Ocular Lens Power Calculation and Optic Biometry...
Visual Field Defects in Glaucoma
Visual Field Defect and Retinal Nerve Fiber Layer Defect in a Case of Optic Nerve Head Drusen...
Current Minimal Invasive Angle Procedures Without Implants for the Treatment of Glaucoma...
Intra Ocular Lens Power Calculation and Optic Biometry...
Automated Perimetry
Carbonic Anhydrase Inhibitors
Visual Field Defect and Retinal Nerve Fiber Layer Defect in a Case of Optic Nerve Head Drusen...
Glokom-Katarakt 2022 , Vol 17 , Num 1
Abstract Free Full Text Similar Articles Mail to Author
The Effect of Gonioscopic Examination Using a Four- Mirror Goniolens on Intraocular Pressure
Raziye Dönmez Gün1, Ayşegül Penbe1
MD, Istanbul Kartal Doctor Lütfi Kırdar City Hospital, Department of Ophthalmology, Istanbul, Turkey DOI : 10.37844/glauc.cat.2022.17.3 Purpose: To evaluate whether gonioscopic examination using a four-mirror goniolens affects intraocular pressure (IOP).

Materials and Methods: This cross-sectional study included one randomly selected eye of 30 patients with primary open-angle glaucoma (POAG) under medical treatment and 27 healthy controls. IOP was measured using a noncontact tonometer (NT), then with a Tono-pen, followed by gonioscopic examination using a four-mirror goniolens (without gel), and finally a second IOP measurement by Tono-pen. We compared NT and Tono-pen measurements obtained before gonioscopic examination, and Tono-pen measurements obtained before and after gonioscopic examination. We evaluated the relationship between the difference in Tono-pen measurements before and after gonioscopy and central corneal thickness (CCT) and palpebral fissure height (PFH).

Results: Mean IOPs in the POAG and control groups were 15.7±3.2 and 16±2.7 mmHg with NT before gonioscopy, 15.3±2.6 and 14.81±2.27 mmHg with Tono-pen before gonioscopy, and 14.9±3.1 and 14.37±2.13 mmHg with Tono-pen after gonioscopy, respectively. In pre-gonioscopic measurements, NT and Tono-pen measurements did not differ significantly in the POAG group (p>0.05), whereas NT measurements were higher than Tono-pen measurements in controls (p<0.05). The mean change in IOP after gonioscopy was ?0.32±1.60 and ?0.44±1.69 mmHg in the POAG and control groups, respectively (p>0.05). There was no difference in CCT between groups (p>0.05), whereas PFH was smaller in controls (9.67±0.73 vs. 10.1±0.55 mm; p<0.05). Change in IOP was not correlated with CCT or PFH in either group (p>0.05).

Conclusions: Our results indicate that gonioscopic examination using a four-mirror goniolens has no significant effect on IOP in open-angle eyes. Keywords : Gonioscopy, Intraocular pressure, Glaucoma